background image

Setup

31

ninb63e1-f (1410)

6.4

Set Up Test Area

Test Area Recommendation

Place fresh air intake on outer wall of building.

Place air intake far away from tracer gas exhaust, cargo bays and other hydrogen
sources.

Do not use compressed air as fresh air supply.
Industrial compressed air often contains varying and substantial amounts of
hydrogen.

Fresh Air Curtain Recommendation

Try to create a laminar flow over the test area.

Curtain should cover the entire test area (test hood or sample point) and extend at
least 0.5 m outside the area.

Air speed in curtain should be rather low, typically 0.1 m/s.

Additional small fan(s) can be set up within the curtain for directional purging of test

chamber etc.

1

Fresh Air Fan

2

Exhaust Fan

3

Test Area

4

Test Building

1

Fan

2

Local Air Jet

3

Test Object

4

Filter

1

3

4

2

Large distance

3

1

2

4