6 control output – INFICON Sensistor ILS500 Leak Detection System User Manual

Page 93

Advertising
background image

Technical Data

93

ninb63e1-f (1410)

14.4.6

Control Output

Purpose:

Electrical tooling interface.

Pin

Signal

Type

Load

Comment

1

+24 VDC

SUPPLY

300 mA

Tooling switch supply (e.g. proximity
switch).

2

TS1

IN

-7 mA

Tooling switch 1.

3

TS2

IN

-7 mA

Tooling switch 2.

4

TS3

IN

-7 mA

Tooling switch 3.

5

TS4

IN

-7 mA

Tooling switch 4.

6

MARKER*

OUT

0.5 A

Marker output. Selectable mark on
REJECT or ACCEPT.

7

COM

GND

-1.0 A

Common GND.

8

COM

GND

-1.0 A

Common GND.

*

MARKER output (Tooling Connector, pin 6) can be used to send a start pulse to
marking equipment such as an engraving machine or a valve controlling a simple
pneumatic stamp. Function and length of pulse is set by the following two
parameters:

Marker Output:

Length of marker output pulse.
Output will go high at end of gas test and stay
high for the given time.

Marker Output High if Leak:

Decides function of marker pulse.
To mark rejected part set to OFF.
To mark accepted part, set to ON.

Connector:

8 pin male Weidmüller, Omnimate BL3.5. Mating screw terminal
included.

Purpose:

External start and stop. Control of optional external valves.

Pin

Signal

Type

Load

Comment

1

+24 VDC

SUPPLY

2.0 A

Start and stop switch and supply.

2

EXTSTART

IN

-7 mA

Start button return (NO contacts) or
contact to +24 VDC.