INFICON Sensistor ILS500 Leak Detection System User Manual

Page 74

Advertising
background image

74

Troubleshooting

ninb63e1-f (1410)

Tested Unit

Tested Feature

Reason for Error

Action

Tracer Gas Fill Valve

Internal leaks

Dirty or worn valve seals.

Replace or clean tracer
gas fill valve.

Leaking pilot valve.

Replace fourth valve from
bottom in pilot valve ramp.

Test Port 2 Valve

Function

Dirty or broken pilot valve. Replace third valve from

bottom in pilot ramp.

Test port 2 valve broken.

Replace valve.

Tracer Gas Fill Valve

Function

Dirty or broken pilot valve. Replace fourth valve from

bottom in pilot ramp.

Tracer gas fill valve
broken.

Replace valve.

Gas Valve Manifold

Leaks to outside

Wrongly assembled gas
valve.

Remove leaking valve.
Clean and grease valve
seal before installing
again. See instructions.

Wrongly installed
connectors/plugs.

Remove leaking unit.

Clean and grease o-ring

Install again.

Units lacking o-ring seal
should be sealed with
Loctite 577 or similar.

Evacuation Valve

Function

Dirty or broken Evacuation
pilot valves.

Replace fourth valve from
bottom in pilot valve ramp.

Lamp

Function

Broken lamp.

Replace lamp.

Send in for repair.

Tooling Valves

Function

Dirty or broken pilot valve. Replace first or second

valve from bottom in pilot
ramp.

Buttons

Function

Broken switch.

Send in for repair.