INFICON Sensistor ILS500 Leak Detection System User Manual

Page 67

Advertising
background image

Troubleshooting

67

ninb63e1-f (1410)

Test Port 2 Valve

Hardware Test

CONTINUE

STOP

Vacuum:

Pressure:

0.00 bar

bar

Test Port 2 Valve

Valve works!

0.00

This step tests Test Port 2 valve for internal leakage by registering pressure rise behind
the valve.

Possible results:

No Internal Leakage

Internal Leakage

Internal leakage can result in:

False blockage test accepts

Tracer Gas Fill Valve

Hardware Test

CONTINUE

STOP

Vacuum:

Pressure:

0.00 bar

bar

Tracer Gas Fill Valve

Valve works!

0.50

This step tests that tracer gas fill valve opens to fill gas. Test will fail if tracer gas feed
pressure is too low. If this is the case, adjust pressure and restart hardware test from
beginning.

Possible results:

Valve works

Faulty!

Malfunction will result in:

Failed gas filling

Notice

This test will fail if both test ports are connected to a test object. Proceed
and then repeat the entire hardware test sequence with both ports
plugged to perform this test step.