12.6.2 Summary Reports

The Summary report provides the method start date and time, the name of the
method file, the name of the tune file, a description of the method, the name of the
data file and the sample description if entered into the Data File Information window
(see

section 12.5 on page 12-5

). A list of compounds identified, the retention times

and the concentrations will also be displayed. The compounds listed will be in order
of their retention time. See

Figure 12-11

.

Figure 12-11 Summary Report

12.6.3 Quantitative Reports

A Quantitative report will be displayed for methods containing a calibration library.
The report will show the method start date and time, the compounds identified, the
predicted retention time, the actual retention time, the area, the concentration and
the limit. If the concentration is above the limit, it will be noted in the report. See

Chapter 15, Method Editor

. The flag column will provide additional information

when necessary. For example, Area too low may be displayed in the Flag column,
if the area of the peak was lower than the method limit for peak area. See

Figure

12-12

.