CMS5000 Operating Manual
Chapter 5
Software Setup
5.1 Software Setup
Prior to running continuous analyses, CMS5000 methods require calibration. This
chapter provides instructions for calibrating the Check Standard method, the Water
Purge method, and the Air Purge method.
Various software parameters will need to be established in order to operate the
CMS5000 and retrieve data. This includes entering startup and input methods,
setting up the FTP server(s) and if desired, sequencing methods.
5.2 Calibrating the Check Standard
1
Run the check standard method, CMS5000 Ck Std. See
for information on running a method.
NOTE: The onboard toluene permeation tube is considered to be equilibrated
when a stable response for toluene is observed. Consecutive runs of
the check standard method may need to be run until the toluene
response is stable.
2
The finished run will look similar to the chromatogram shown in
.
Figure 5-1 Check Standard Method Run
3
When the method has finished running, open CMS IQ on the PC.
Figure 5-2 CMS IQ