CMS5000 Operating Manual

Chapter 2

Specifications

2.1 Gas Chromatograph Specifications

Column . . . . . . . . . . . . . . . . . . . . . . . 100% dimethylpolysiloxane, 0.32 mm id,

4.0

µ

m df or equivalent

Valves . . . . . . . . . . . . . . . . . . . . . . . . Stainless Steel body/Teflon diaphragm

Temperature Programmable
Column Module . . . . . . . . . . . . . . . . 55-200

°

C

Carrier Gas. . . . . . . . . . . . . . . . . . . . Argon 99.9999% @ 414-689 kPa

(60-100 psi)

Variable Column Pressure Control . . Regulator pre-set to 90 psi

Sample Inlet . . . . . . . . . . . . . . . . . . . Water Vessel with purge-and-trap SituProbe

Air Sampling Tube

2.2 MAID - Micro Argon Ionization Detector

Temperature . . . . . . . . . . . . . . . . . . . Maximum 110

°

C

Sensitivity . . . . . . . . . . . . . . . . . . . . . 0.5 ppb Benzene in water, s/n > 200:1

0.5 ppb MTBE in water, s/n > 15:1

Detection Limit . . . . . . . . . . . . . . . . . PPB- PPT in air

Dynamic Range . . . . . . . . . . . . . . . . 3 decades

Stability . . . . . . . . . . . . . . . . . . . . . . . 5 runs of 5 ppb Benzene in air and water:

RSD calculated <15%

Radioactive Source . . . . . . . . . . . . . 2.4 mCi Ni-63

2.3 Communication

Integration . . . . . . . . . . . . . . . . . . . . TCP/IP based

USB for local diagnostics
I/O Relay Contacts

FTP. . . . . . . . . . . . . . . . . . . . . . . . . . Configurable for automated data upload

Storage/Archival . . . . . . . . . . . . . . . . 16 GB Flash Drive

System Status . . . . . . . . . . . . . . . . . Status table for system operating changes

Wireless Connectivity . . . . . . . . . . . . 802.11 B/G

Optional Touch Screen . . . . . . . . . . . 6.5" VGA color display with touch screen