16.2 Using the Calibrate Function

NOTE: For detailed information regarding the calibration functions, see

section

16.3, Glossary of Terms in the Calibrate Window, on page 16-17

.

WARNING

Wear appropriate Personal Protection Equipment (PPE)
as advised in the MSDS of the standard(s) being used.

1

Various standard concentrations must be prepared for the desired analytes.

2

After the standard(s) have been prepared, each concentration will be run
separately on the CMS5000 using the desired method. Each standard will have
its own separate run. The method used for the CMS5000 runs can be a default
method or a custom method created with the Method Editor.

NOTE: All method development must be made prior to running the standards

on the CMS5000. Refer to

Chapter 15, Method Editor

, for additional

information on method development.

3

Enter the concentration of the standard and a description on the Data File
Information
page during each sample run. See

Figure 16-1

. To access this

window, refer to

section 12.4, Data Review Toolbar, on page 12-4

.

Figure 16-1 Data File Information Window