CMS5000 Operating Manual
15
Enter the concentration of each analyte in the highlighted field. If the
concentrations were entered into the Data File Information window, this field
will automatically populate. Refer to
. See
Figure 16-13 Entering Analyte Concentration
16
Enter the Water Beta value when calibrating water methods (check standard
and air methods do not need a Water Beta value). Published reference
materials are available to obtain these values. See
for a definition.
See
.
Figure 16-14 Water Beta Value
17
Repeat
through
for all remaining compounds.
18
To save the new library, select the Save Library button in the Method box. See
Figure 16-15 Save Library Button
19
Fill in the Enter Library Name and Enter Creator Name fields as desired.
Click OK. See
Figure 16-16 Save Library