INFICON CMS5000 Monitoring System User Manual

Page 49

Advertising
background image

5 - 5

IP

N 07

4-

50

8-

P1

C

CMS5000 Operating Manual

9

Verify that toluene is the only component in the library. See

Figure 5-10

.

Figure 5-10 Is Check Std Library Box

NOTE: If the Methods Differ message is displayed, verify that the differing

parameters are acceptable and click Yes. If the parameters are not
acceptable, run a calibration sample with the exact method selected in
the Method box and use that data file. See

Figure 5-11

.

Figure 5-11 Methods Differ Message

10

A message will be displayed stating that the units between the data files are
inconsistent. This message is displayed because units for the data files have
yet to be selected. Click OK to continue. See

Figure 5-12

.

Figure 5-12 Inconsistent Data Files