8 method sequence – INFICON CMS5000 Monitoring System User Manual

Page 69

Advertising
background image

5 - 25

IP

N 07

4-

50

8-

P1

C

CMS5000 Operating Manual

5

The method will be cleared from the selected input method line.
See

Figure 5-53

.

Figure 5-53 Method Cleared

5.8 Method Sequence

A series of methods can be selected to run at timed intervals, which includes
consecutive runs. Once the series has been started, it can run unattended and
unprompted until the sequence has finished. Follow the instructions below to
create a method sequence.

1

From the CMS IQ Setup screen, double-click the Method Editor icon.
See

Figure 5-54

.

Figure 5-54 Method Editor Icon

2

The Edit Method window will be displayed. Click on Method Sequence.
See

Figure 5-55

.

Figure 5-55 Edit Method Window