CMS5000 Operating Manual
24
If Global is selected, enter the concentration of the standard in the data file in
the Conc/Factor column. See
Figure 16-21 Adding Concentrations using the Global Setting
25
When Global is selected, the Concentration Column in the center section will
automatically be set to 1. This column does not require further modification in
order to properly calibrate the standards. See
Figure 16-22 Concentration Column Using the Global Setting
26
If Analyte is selected, enter the individual concentration of each analyte into
the Concentration Column in the center section. See
Figure 16-23 Entering Concentrations using the Analyte Setting