6.5 STAT Option

The STAT option displays temperature information and carrier gas pressure. This
includes the current temperatures and the setpoint temperatures of the Regulator
Heater
, Check Standard, Valve Oven, Column, Detector (MAID), and the Argon
Supply
pressure. The state of the heaters will be shown, as well, with ctrl
indicating that the heaters are within 2 degrees of their setpoint. The temperature
of the water sample and the temperature of the card cage, which holds the boards
of the unit, are located beneath the temperatures of the heated components. The
pressure of the argon carrier gas is the last listing on the screen. See

Figure 6-4

.

NOTE: If the argon pressure is 380 kPa or less, a low-pressure warning will

appear. When the pressure drops to 350 kPa, a low pressure error
message will be displayed.

Figure 6-4 STAT Option

6.6 FIRM Option

The FIRM option displays the firmware version for the gas chromatograph (GC)
and the front panel (FP). See

Figure 6-5

.

Figure 6-5 FIRM Option