INFICON CMS5000 Monitoring System User Manual

Page 242

Advertising
background image

16 - 12

IP

N 07

4-

50

8-

P1

C

CMS5000 Operating Manual

27

Enter the factor of the standard in the Conc/Factor column in the Data Files
section on the left side of the screen. This number will multiply each
Concentration entered into the library template. For example, 1 µL of a
specific compound in a standard has a concentration of 1 ppb. If the standard
was prepared using 5 µL, the factor of the standard would be 5, because five
times the amount of original concentration was used. See

Figure 16-24

NOTE: For more details about Concentration Reference or Conc/Factor, see

section 16.3, Glossary of Terms in the Calibrate Window

.

Figure 16-24 Conc/Factor

Advertising