16.3.7 Calibrate Display

Figure 16-41 Calibrate Display

Check Std Auto Cal Library . . . . . . For use with the Check Standard Method

only. When selected, the Check Standard
Method can be overwritten to create a new
check standard auto calibration.

# . . . . . . . . . . . . . . . . . . . . . . . . . . . . The assigned analyte number in the library.

Data Ref . . . . . . . . . . . . . . . . . . . . . . Displays the data file reference number.

Compound . . . . . . . . . . . . . . . . . . . Displays the compound name that is

assigned by the user for the analyte.

CAS #. . . . . . . . . . . . . . . . . . . . . . . . Chemical Abstracts Service number. This is

a unique number assigned to each chemical
that is accepted worldwide.

Ret. Time . . . . . . . . . . . . . . . . . . . . . The retention time for the analyte.

Area . . . . . . . . . . . . . . . . . . . . . . . . . Displays the integrated analyte area.

Standard . . . . . . . . . . . . . . . . . . . . . Designates the compound as an analyte or

an internal standard. Analyte is more
frequently selected, as most CMS5000
methods do not use internal standards.

Conc . . . . . . . . . . . . . . . . . . . . . . . . Displays the concentration of the analyte in

the displayed file.

Water Beta. . . . . . . . . . . . . . . . . . . . Numerical value that corrects for the

changes in physical properties of analytes
due to fluctuations in water temperature.
Refer to published reference material to
obtain these values.

Water Temp . . . . . . . . . . . . . . . . . . . The temperature of the water being analyzed

by the CMS5000.