CMS5000 Operating Manual

Events can be deleted from the template. Highlight the desired cell in the grid and
press the Delete key on the PC keyboard to remove a row.

Events can also be inserted to the template. Highlight the desired cell in the grid
and press the Insert key on the PC keyboard to insert a row.

NOTE: Rows cannot be inserted after the Standby event.

15.5.2 GC Temperature Profile

GC Temperature Profiles specify the column temperature, ramp rate and hold
settings for the CMS5000 Method. See

Figure 15-15

.

Figure 15-15 GC Temperature Profiles

Adjustments to the Hold Time, Ramp Rate and Temp (Temperature) columns will
automatically update dependent parameters. For example, increasing the Temp
will increase the Ramp Time and increasing the Hold Time will adjust the Start
time of the next parameter. See

Figure 15-16

.

Figure 15-16 Adjusting Temperature

NOTE: A maximum of four lines is allowed in this section.