6 search page – INFICON CMS5000 Monitoring System User Manual
Page 219
15 - 13
IP
N 07
4-
50
8-
P1
C
CMS5000 Operating Manual
15.6 Search Page
The Search page sets the necessary parameters to qualify and quantify data. See
Figure 15-18
. To quantify data, a calibration library must be created. See
16, Creating Calibration Libraries
for instructions on creating a calibration library.
Figure 15-18 Method Editor Search Page
There are two choices in the Search Mode drop-down menu.
No Search . . . . . . . . . . . . . . . . . . . . If this option is selected, a library search will
not be conducted and a report will not be
displayed on the front panel at the end of a
run. If there is no library present, this is the
only option available.
Quantitative. . . . . . . . . . . . . . . . . . . Generates a Quantitative (Quant) report at
the end of a run by referencing the
designated library