Chapter 3 menus – INFICON SQC-310 Thin Film Deposition Controller User Manual

Page 40

Advertising
background image

Chapter 3

Menus

3-2

3.1 Main Screen, Menu 1

The table below describes the function of each SoftKey on Main Screen, Menu 1.

Next
Menu

Sequences through each of the three Main Screen menus.

Quick
Edit

Displays the Quick Setup Menu of commonly changed process
values. If this key is not visible, the active process has no
layers defined.

Auto /
Manual

Toggles between Auto and Manual power control. When
Auto/Manual is shown, output power is set by the SQC-300 to
achieve the programmed deposition rate. When Manual/Auto
is shown, the control knob sets the output power.

Zero

Zeros the thickness reading. Useful for resetting or extending
the current deposition layer.

Next
Layer

Sequences through each process layer. Use this key to start
or restart the process at any layer. Only visible when the
process is stopped.

Start
Layer

Each layer in a process can be defined as Auto Start or Manual
Start. Auto Start layers begin immediately on completion of the
previous layer. Manual start layers wait for the operator to
press Start Layer. Only visible when waiting to start a Manual
Start layer.

Start/Stop

Starts or halts the current process. Sets all outputs to zero.

Advertising