Chapter 5 communications – INFICON SQC-310 Thin Film Deposition Controller User Manual

Page 78

Advertising
background image

Chapter 5

Communications

5-10

Get Sensor Thickness

Command:

N

Description: Returns the thickness reading for the requested sensor.

Parameters: Sensor Number

Example:

To Get Sensor 1 thickness, send: N1

The return string is of the form: 1.000

Get Output Thickness

Command:

O

Description: Returns the average thickness of all sensors assigned to the requested
output.

Parameters: Output Number

Example:

To Get Output 1 thickness send: O1

The return string is of the form: 1.000

Get Sensor Frequency

Command:

P

Description: Returns the frequency of the requested sensor.

Parameters: Sensor Number

Example:

To Get Sensor 1 frequency, send: P1

The return string is of the form: 5543210.0