Chapter 5
Communications
5-9
Get Readings
Command:
K
Description: Returns the phase time and sensor or output readings for all installed
channels.
Parameters:
1 = Output Readings
2 = Sensor Readings
Example:
To Get Output readings, send: K1
The return string is of the form:
Time Rate1 Dev1 Thick1 Power1 Rate2 Dev2 Thick 2 etc.
To Get Sensor readings, send: K2
The return string is of the form:
Time Rate1 Thick1 Frequency1 Rate2 Thick 2 etc.
Get Sensor Rate
Command:
L
Description: Returns the sensor rate for the requested sensor.
Parameters: Sensor Number
Example:
To Get Sensor 1 rate, send: L1
The return string is of the form: 1.00
Get Output Rate
Command:
M
Description: Returns the average rate of all sensors assigned to the requested output.
Parameters: Output Number
Example:
To Get Output 1 rate, send: M1
The return string is of the form: 1.00