Chapter 5 communications – INFICON SQC-310 Thin Film Deposition Controller User Manual

Page 75

Advertising
background image

Chapter 5

Communications

5-7

Delete All Layers

Command:

E

Description: Deletes all 250 layers! Most often used to clear the entire inreument in
preparation for downloading a new list.

Parameters: None

Get/Set Material Parameters

Command:

F

Description: Gets/Sets the parameters of the 100 stored materials.

Parameters:

1.

Material Name

3.

Zfactor

2.

Density

Example:

To Get Material 1 name, send: F1? 1

The return string would be of the form: Aluminum

To Set the parameter you would send: F1 1,AnyName

Note: You can only get/set one material parameter at a time.

Get/Set Input Parameters

Command:

G

Description: Gets/Sets the logical function of each of the 16 digital inputs.

Parameters:

1

Start Process

32

Soak Hold

2

Abort Process

33

Zero Thickness

3

Start Layer

34

Zero Time

4

Stop Layer

35

Out1 Pocket Ready

5

Start Next Layer

36

Out2 Pocket Ready

6

Force Final Thick

37

Out3 Pocket Ready

7-31

Start Process 1-25

38

Out4 Pocket Ready

39

Not Used

Example:

To Get Input 1 and Input 2 functions, send: G? 1 2