Chapter 4 maintenance – INFICON SQC-310 Thin Film Deposition Controller User Manual

Page 67

Advertising
background image

Chapter 4

Maintenance

4-3

4.3 Option Card Install ation

The option card adds two additional sensor inputs, two control outputs, and 8 digital I/O
signals to the SQC-300. To install the option card follow these steps.

1. Unplug the SQC-300 from the main power!
2. Remove (and save) the eight screws securing the top cover. Remove the cover.
2. Remove the three screws securing the option cover plate to the back panel.
3. Remove the screws from the two 5/8” standoffs near the rear of the main PCB.
4. Plug the short ribbon cable into the connector on the main PCB.
5. Slide the option card PCB into the back panel.
6. Secure the option card PCB to the main PCB with two 2” screws and lockwashers.
7. Screw the four BNC hex nuts to the back panel.
8. Screw the two jack screws and washers into the back panel I/O connector.
9. Replace the cover and attach with the eight screws saved in step 1.
10. At turn on the SQC-300 will now display four channels on the main display.

4.4 Half Rack Adapter Installation

The Half Rack Adapter option (PN 900-016) adapts the SQC-310 model to many other
5 1/4” high half rack instruments. The Half Rack Adapter option (PN 900-017) adapts
the SQC-330 model to many other 3 1/2” high half rack instruments.

The option consists of two standard rack mount ears and an adapter bracket. Attach
one of the standard rack mount ears to the SQC-300 with the 10-32 flat head screws
supplied. Attach the adapter bracket to the other side of the SQC-300.

Place the other instrument so that its rack mount ear matches the two threaded 10-32
holes on the adapter bracket. Attach the other instrument to the adapter bracket with
two standard 10-32 rack mount screws (not supplied).

Advertising