Input/output characteristics, Source control voltage output, Sensor input – INFICON MDC-360C Thin Film Deposition Controller User Manual

Page 103: Discrete outputs, Discrete inputs, Input/output characteristics -1, Source, Control, Voltage, Output -1

Advertising
background image

MDC-360C DEPOSITION CONTROLLER

INPUT/OUTPUT CHARACTERISTICS

7-1

7. INPUT/OUTPUT

CHARACTERISTICS

The following section describes the electrical characteristics of the MDC-360C
inputs and outputs. All outputs are updated and inputs are sampled every 100
msec. In order to insure immunity to transients, inputs are not considered to have
changed until the same input state is obtained on two successive input samples.

For this reason all input signals must have a minimum duration of at least 0.2 sec.
Input signals lasting less than 100 msec. will be ignored while signals lasting
between 100 and 200 msec. may or may not be recognized.

7.1

SOURCE CONTROL VOLTAGE OUTPUT

For maximum noise immunity, each two-terminal control voltage output pair is
isolated from controller ground. Either terminal can be grounded within the user
system, so the output can provide either a negative or positive output voltage
range. In the event that the receiving equipment has an isolated input, one of the
two lines should be grounded to avoid excessive voltage buildup on the otherwise
isolated circuitry.

The voltage output range can be programmed (see Source Setup) for 2.5, 5.0 or 10
volts full scale. The output impedance is nominally 100 ohms. The outputs are
short circuit protected with short circuit current limited to between 20 and 40
milliamps, though the outputs should not be short-circuited for long periods. The
schematic appears in Figure 7-5.

CAUTION: To avoid damage to the instrument, do NOT allow long term
shorting of any of the source output which may cause excessive temperature
rise in the isolated power supply.

7.2 SENSOR

INPUT

The sensor oscillator is connected through a single coaxial cable. Sensor ground
is common with the MDC-360C ground. Power to the sensor oscillator is carried
on the center conductor of the coaxial cable. Power is supplied from the MDC-
360C internal 5 volt supply through a 50 ohm resistor which accomplishes the
dual function of properly terminating the 50 ohm coaxial cable and providing
short circuit protection. The sensor buffer circuit is shown schematically in
Figure 7-4.

7.3 DISCRETE

OUTPUTS

Each Discrete Output is an isolated, independent, normally open relay output
connected to one pin pair on the output connector. See Table 8-4 for pin signal
assignments.

7.4 DISCRETE

INPUTS

The Input circuit for the Passive I/O card is shown in Figure 7-1. The Passive
inputs are activated by shorting the input’s pins together. The inputs are

Advertising