MDC-360C DEPOSITION CONTROLLER

PROGRAMMING AND CONTROLLER SETUP

4-42

This parameter sets the upper voltage range for the source control output. The
lower voltage range is always 0. For example, selecting 10 for this parameter sets
the source control voltage range from 0 to 10 volts.

4.3.3.7 EDIT DAC SETUP

Selecting Edit DAC Setup from the Edit System Setup menu will present the
DAC Setup screen that allows selection of the parameter and its signal range for
each of the two DACs.

Figure 4-19 DAC Setup Screen

1. DAC Output (Rate, Rate Dev., Power, Thickness)

One of four system control parameters is chosen for the DAC output. The default
setting is Rate for DAC #1 and Rate Deviation for DAC #2.

2. DAC Scale (2-digit, 3-digit)

Either the two least significant, or the three least significant, digits of the chosen
control parameter are used to represent full scale for the DAC output.

4.3.3.8 EDIT UTILITY SETUP

Selecting the Edit Utility Setup from the Edit System Setup menu will present the
Utility Setup screen. Figure 4-20 shows the first page of this screen. All
parameters are described below.

DAC Setup


DAC Output #1

Rate

-----

DAC Scale #1 3 Digit
DAC Output #2 Power
DAC Scale #2 3 Digit