Program actions, Program actions -34, Figure 4-14 action selection screen -34 – INFICON MDC-360C Thin Film Deposition Controller User Manual

Page 74: Figure 4-15 program action screen -34

Advertising
background image

MDC-360C DEPOSITION CONTROLLER

PROGRAMMING AND CONTROLLER SETUP

4-34

while it is in the Predeposit Hold state then one second after it enters the deposit
state, the "Ion Gun Power" output will go true for 180 seconds or three minutes.

4.3.3.4 PROGRAM ACTIONS

The MDC-360C provides for 16 internal user programmable actions. Internal
actions are used to provide special functions at the true evaluation of a condition
string. These functions may be such things as terminating a deposit on an input
from an optical monitor. Or, sounding an alarm when certain events are true.

To program an action, first select the desired action from the list of 16
programmable actions displayed in the Actions screen.

Figure 4-14 Action Selection Screen


Once you have selected the required action, pressing the Right-arrow key will
present the screen which permits programming of the action details, and this
procedure is similar to the one used for programming discrete outputs.

Figure 4-15 Program Action Screen

In this screen you select the predefined action you would like to take and the
associated conditions. To specify an action, move the markers onto the action
name field and press the Right-arrow key. This will present the Select Defined
Action screen.

Select Action 01

Hold In State

-----

02 Step From State

03 Sound Attention

04 Sound Alert

05 Sound Alarm

06 No Action

07 No Action

- View/Edit 08 No Action

’

Action Name

Hold In State

-----

Conditions





Press > to select action

Advertising