MDC-360C DEPOSITION CONTROLLER

BENCH CHECKOUT & INSPECTION

3-4

3.3.3 PROCESS

PARAMETERS

Process Name

Layer No.

Thickness

Material

Sample 1

0.400

Cr

2

1.050

Au

3.4 SIMULATE

OPERATION

Testing the MDC-360C is best accomplished by checking its operation in the
Simulate mode. This mode can be selected by using the programming Main
Menu, Edit System Setup, Edit Utility Setup, to select Simulate mode ON, then
use Start to select and run a process in Simulate mode.

The Simulate mode is identical to the Normal mode except that the sensor inputs
are simulated. For this reason, entry to the Simulate mode will extinguish the
Crystal Failure message if it is flashing. No other difference between the
Simulate mode and the Normal mode occurs until entry to the Deposit State.

3.5 MANUAL

OPERATION

Manual Mode is selected by depressing the Manual key. The LED behind the key
will light up indicating the controller is in Manual mode.

The Manual Mode is identical to the normal mode in all respects except that
source power for the active source is controlled only through the Remote Power
Handset.

The Remote Power Handset has three push buttons, see Figure 3-1. Without any
of the buttons depressed, the output power is maintained at its last value.
Depressing the “PWR UP” button will increase the power, depressing the “PWR
DN” button will decrease the power and depressing the “ABORT” button will put
the controller into the Abort mode.

The Abort Mode is active whether or not the MDC-360C is in Manual Mode and
therefore can be used as a remote “panic button”.

The minimum increment by which the power is increased or decreased is 0.1%.

3.6

INSTALLING OPTION BOARDS

Option boards are most easily installed while the MDC-360C is on the bench.
Figure 8-10 shows the location of the various option boards. Also, they are
clearly marked on the rear panel.

All Dual Source-Sensor boards are identical, as are all Discrete I/O boards. The
input-output configuration of these boards is defined by the position into which
they are installed. One exception for the Discrete I/O boards is that the jumper J2
on the board installed in the Discrete I/O-2 position has to be connected. This is
required so the controller will acknowledge the second Discrete I/O board.

A Source-Sensor board plugged into the second position will provide sensor
inputs numbers 3 & 4 and source outputs numbers 3 & 4. A Source-Sensor board

Advertising