System setup, Edit display setup, System setup -25 – INFICON MDC-360C Thin Film Deposition Controller User Manual

Page 65: Edit display setup -25, Figure 4-7 system setup menu screen -25, Figure 4-8 display setup screen -25

Advertising
background image

MDC-360C DEPOSITION CONTROLLER

PROGRAMMING AND CONTROLLER SETUP

4-25

4.3.3 SYSTEM

SETUP

Choosing the Edit System Setup option from the Main Menu screen will present
the System Setup Menu options as shown in Figure 4-7. These options allow for
setting up the controller to interface with the vacuum system and are described
below.

Figure 4-7 System Setup Menu screen

4.3.3.1 EDIT DISPLAY SETUP

Selecting Edit Display Setup will present the Display Setup screen.

Figure 4-8 Display Setup Screen

1. Pause On Layer Complete (Yes or No)

This parameter determines whether the controller will pause between layers. If
this parameter is set to Yes then the controller will stop on layer complete and
wait for a Start key press from the operator. If this parameter is set to No then the
controller will automatically increment to the next layer.

2. Display Negatives (Enabled, Disabled)

This parameter defines whether the MDC-360C will display negative rates and
thickness or not. If set to disable, the MDC-360C will hold negative values at
zero. The default for this parameter is Disabled.

3. Thickness Graph Scale (2-digit, 3-digit)

System Setup

Edit Display Setup

Program Inputs
Program Outputs
Program Actions
Edit Sensor Setup
Edit Source Setup
Edit DAC Setup
Edit Utility Setup

Display Setup

Pause On Layer Complete

Yes

Display Negatives Enabled
Thickness Graph Scale 3-digit
Sensor Status Thickness
Time Display Estimated Layer
Rate Graph Disabled

Power Graph Enabled

’