Leica CM1950 – Cryostat

2.

Safety and Design

To lock the handwheel, press the lever (1) outward. Continue turning

the handwheel slowly until the grip is in the upper or lower position and

the handwheel is locked. Press the lever fully outward; gently rock the

handwheel back and forth until the locking mechanism clicks into place

noticeably.
To release the handwheel, press the lever (2) on the handwheel toward the

cryostat housing.

Centering of the handwheel (motorized instruments only)
Pull out the handwheel's handle and position it in the middle of the hand-

wheel. The handle automatically engages in this position.

12 o'clock

position

6 o'clock

position

1

2

Fig. 7

Knife guard

Prior to making modifications to the knife and specimen, changing the specimen or knife,

or taking a break, always lock the handwheel and cover the cutting edge with the knife

guard!

The CE, CN and CN-Z knife holders feature knife guards; the glass anti-roll plate of the CE knife

holder also serves as a knife guard.

Rotate the handwheel only if the refrigeration system is on and the

cryochamber is cold.

An important safety device on the cryostat is the centering of the

handwheel for motorized instruments.

Always lock the handwheel prior to making modifications to the

knife or specimen, changing the specimen, or taking a break!

2.6 Securing/locking the handwheel

Fig. 9

Fig. 8