Optional accessories – Leica Biosystems CM1950 User Manual

Page 65

Advertising
background image

65

Leica CM1950 – Cryostat

Knife holder CNZ with anti-roll guide
• Pressure plate permits the full length of the

knife to be utilized.

• Use hard metal and steel knives here.

11.

Optional Accessories

Knife guard/lateral movement for knife

holder CN

The knife guard (18) is fixed and integrated into

the clamping jaws. The knife guard has handles

(19) that allow it to be moved. The knife guard

is adequate for knives up to 22 cm in length.

Always cover exposed parts of the knife blade

after sectioning.
The anti-roll guide system can be moved sideways

(only for the 84 mm variant). To better find the mid

position, a groove (17) is provided in the shaft (16).

19

19

18

16

17

3a

Fig. 76

After removing the knife from the knife holder,

place it safely into the knife case. NEVER place

it onto the work surface next to the instrument!

The height of resharpened knives must

be adjusted using knurled screws (3a)

(approx. 1 mm under the edge of the

clamping jaw).

Ensure that the knife is adjusted in par-

allel vertically.

Fig. 74

Fig. 75

The safety gloves included in the stan-

dard scope of delivery must be worn

when inserting/ejecting the knife!

Advertising