1 locking/unlocking the handwheel, Installation – Leica Biosystems CM1950 User Manual

Page 25

Advertising
background image

25

Leica CM1950 – Cryostat

To lock the handwheel, move its handle to the 12 or 6 o'clock position. Press

the lever (1) fully outward; gently rock the handwheel back and forth until

the locking mechanism clicks into place noticeably.
To release the handwheel, press the lever (2) on the handwheel toward the

cryostat housing.

12 o'clock

position

6 o'clock

position

1

2

6.3.2 Installing the footswitch dummy (instruments with cutting motor)

• The footswitch dummy must be installed on the outer right side of the

instrument (

see Chap. 5 "General Overview"

) if no footswitch (optional)

is used.

If the red LED in the E-STOP field in control panel field 3 is illuminated, either:
• emergency-stop function active, or
• Footswitch dummy (opt. foot switch) not connected or incorrectly

connected.

Fig. 18

To do so, pull out the handwheel's handle and position it in the middle of the

handwheel. The handle automatically engages in this position.

Centering the Handwheel (Optional)

6.

Installation

Rotate the handwheel only if the refrigeration system is on and the

cryochamber is cold.

6.3.1 Locking/unlocking the handwheel

Always lock the handwheel prior to making modifications to the

knife or specimen, changing the specimen, or taking a break!

An important safety device on the cryostat is the feature for center-

ing the handwheel in motorized sectioning mode.

Fig. 15

Fig. 16

Fig. 17