3 configuring the parameters, Configuring the parameters, Working with the instrument – Leica Biosystems CM1950 User Manual

Page 40

Advertising
background image

40

Instructions for Use V 1.5 RevF - 08/2014

8.

Working with the Instrument

• Turn the lamp on or off.

• This button activates or deactivates the manual defrost of the cooling

chamber, specimen head or freeze shelf. (For more detailed handling

instructions, refer to the

chapter on "Working with the Instrument"

in

these Instructions for Use.)

• Pressing the key button for approx. 5 sec. locks the entire keyboard (the

LEDs in the clock go out).

• Pressing the key button briefly, then the "-" key in the specimen head

control panel field, switches off the specimen head.

• Pressing the key button briefly, then the "+" key in the specimen head

control panel field, switches the specimen head back on.

Programming the temperature of the cryochamber
• The temperature of the cryochamber is set and indicated on the control

panel marked with the cryostat symbol.

The actual temperature is the standard indication.
Briefly pressing the "+" or "-" button displays the target temperature.
Set the desired value via the "+" / "-" buttons. Pushing the "+" or "-" button

for more than 1 sec. increases or decreases the chamber temperature

continuously.

The actual value will be indicated 5 seconds after finishing the program-

ming.

Fig. 42

The instrument must be switched on at least 5 hours before the planned use.

8.3 Configuring the parameters

Refer to

Chap. 10 "Temperature selection chart"

for a table with guide values. The tem-

perature values given there are based on experience, but are intended solely as guide

values, as any tissue may require particular adjustments.