Instrument controls – Leica Biosystems CM1950 User Manual

Page 33

Advertising
background image

33

Leica CM1950 – Cryostat

7.

Instrument Controls

Display of actual and target specimen

head temperatures

Display of actual and target chamber

temperatures

Display of real time, defrost time

and error messages

Max-Cool button to

select the maximum

low temperature of

the specimen head

directly (-50 °C)

+/- buttons to

select speci-

men head

temperature

+/- buttons

to select

chamber

temperature

+/- buttons to

select defrost

time

+/- buttons to

select real

time

Press the key button to

lock/unlock the entire

keypad.

(

Refer to following page

for

activation of the specimen

head)

"Melting Snow-

flake" button to

activate manual

defrost

ON/OFF button of

illumination

7.1 Control panel fields on the CM1950 – Control panel field 1

Button to

enable/disable

the Peltier

element

UVC

disinfection,

(short time 30 min,

long time 180 min)

Button to

enable/disable

extraction

system

Extraction

force intensity

selectable

from 1-5 (see

following

page for more

information)

EMERGENCY STOP switch to the right of control panel field 1 (motorized

instruments only)

For danger situations during motorized sectioning.

Immediate stop of the sectioning process – motor stops – LED in E-Stop is

illuminated in red.

• Turning in direction of arrow cancels the stop – LED in E-Stop goes out.

• Select single stroke (Single) or continuous stroke (Cont.) operating mode

again.

Fig. 33

Fig. 34