Optional accessories – Leica Biosystems CM1950 User Manual

Page 69

Advertising
background image

69

Leica CM1950 – Cryostat

Footswitch,
for use with motorized instruments only.
The foot switch can be used to control the motorized sectioning process. It

also has a function that is similar to the emergency stop function.

11.

Optional Accessories

• Using the CUT MODE button, select the desired operating mode CONT

or SINGLE on the control panel (

Fig. 38

, page 37).

CONT (continuous stroke) operating mode
• Press the foot switch once briefly to start motorized sectioning.

• Press the foot switch again to stop it.

The specimen then stops in the end position.
SINGLE (single stroke) operating mode
• Press the foot switch once briefly to start motorized sectioning. After

each section, the specimen stops automatically in the end position.

How to activate the emergency stop function
• Press the foot switch strongly to activate the emergency stop function.

Sectioning stops immediately.

The red LED in the E-STOP field on the instrument (

Fig. 38

) is lit up as long

as the foot switch remains depressed.

• To resume the sectioning process, select the sectioning type (CONT or

SINGLE) and restart the system using the foot switch.

If the foot switch remains pressed for longer than half a second, the

specimen stops in the next upper end position.

Caution!

All control panel functions and all buttons on the instrument remain

active along with the foot switch.

Fig. 86