3 assembling the handwheel, Assembling the handwheel, Installation – Leica Biosystems CM1950 User Manual
Page 24
24
Instructions for Use V 1.5 RevF - 08/2014
• Insert the pin (1) of the handwheel shaft into
the hole (2) of the handwheel.
• Tighten the screw (4) using the size 6 Allen
key.
• Place the protective cap on the screw (4).
To dismount, proceed in reverse order.
1
2
4
6.3 Assembling the handwheel
Transport with a fork lift
Fig. 13, Total width with handwheel
6. Installation
• The instrument can be transported with a fork lift.
To ensure a safe transport with a fork lift 3 people are
required: one operating the fork lift, and the other 2
holding the instrument on either side to prevent it from
sliding down.
• At the installation location, unscrew the adjusting feet (1)
(
see Fig. 12
) using the open-end wrench (13 mm). This is
absolutely necessary for the instrument to stand stably.
Rotate the handwheel only if the refrig-
eration system is on and the cryocham-
ber is cold.
Fig. 14