3 assembling the handwheel, Assembling the handwheel, Installation – Leica Biosystems CM1950 User Manual

Page 24

Advertising
background image

24

Instructions for Use V 1.5 RevF - 08/2014

• Insert the pin (1) of the handwheel shaft into

the hole (2) of the handwheel.

• Tighten the screw (4) using the size 6 Allen

key.

• Place the protective cap on the screw (4).
To dismount, proceed in reverse order.

1

2

4

6.3 Assembling the handwheel

Transport with a fork lift

Fig. 13, Total width with handwheel

6. Installation

• The instrument can be transported with a fork lift.

To ensure a safe transport with a fork lift 3 people are

required: one operating the fork lift, and the other 2

holding the instrument on either side to prevent it from

sliding down.

• At the installation location, unscrew the adjusting feet (1)

(

see Fig. 12

) using the open-end wrench (13 mm). This is

absolutely necessary for the instrument to stand stably.

Rotate the handwheel only if the refrig-

eration system is on and the cryocham-

ber is cold.

Fig. 14