Maintenance and cleaning – Leica Biosystems CM1950 User Manual

Page 72

Advertising
background image

72

Instructions for Use V 1.5 RevF - 08/2014

12. Maintenance and Cleaning

Basically, we recommend UV disinfection (for application,

see Chap. 7

"Instrument Controls" - Disinfection

).

For easy-to-use spray disinfection we recommend Leica Cryofect. (Cryofect

is not available in all countries!)

The cryostat has to be disinfected after each daily use.

Emptying the condensate bottle

Check the fill level of the condensate bottle (1) visible in the front panel of

the instrument at regular intervals.
• The bottle collects the condensate that accumulates during defrosting.

1

Fig. 90

In case of visible pollution (such as dust), clean

the air inlet opening (

Fig. 10

, p. 20) of the con-

denser on the bottom right-hand side of the

instrument using a brush, broom or extraction

cleaner in the direction of the louvers.

Be exceptionally careful when clean-

ing the louvers as they have sharp

edges and can cause cutting injuries if

cleaned improperly.

Do not turn the instrument on before the cryochamber is completely dry! Frost formation!

The front panel and the slit cover of the microtome must be completely dry before turning

on the instrument!

All parts that have been removed from the cold cryostat must be dried thoroughly before

they are put back into the chamber.

It is not necessary to oil the parts, such as the T-piece on the micro-

tome base plate, clamping lever etc.

Comply with the instructions for use!

The glass anti-roll plate can remain in place during disinfection.

Dispose of the contents of the bottle in accordance with laboratory

regulations.

Advertising