Troubleshooting – Leica Biosystems CM1950 User Manual

Page 53

Advertising
background image

Leica CM1950 – Cryostat

53

9.

Troubleshooting

Problem

Cause

Remedy

Blunt knife

Inappropriate section thickness.

Incorrect clearance angle.

Microtome not dried thoroughly

enough.

Dried specimen.

Use different part of the knife.

Select correct section thickness.

Set correct angle.

Dry microtome.

Prepare new specimen.

Tissue sticks to the anti-roll

plate

Anti-roll plate is too warm or

incorrectly positioned.

Fat on the corner or edge of anti-roll

plate.

Anti-roll plate not correctly fixed.

Rust on the knife.

Cool down anti-roll plate, or

reposition correctly.

Remove fat from anti-roll plate.

Fix correctly.

Remove rust.

Flattened sections curl

up when anti-roll plate is

folded up

Anti-roll plate too warm.

Cool down anti-roll plate.

Sections tear or separate

Temperature too low for the tissue

cut.

Blunt part, dirt, dust, frost or rust on

the knife.

Top edge of the anti-roll plate

damaged.

Hard particles in the tissue.

Dirt on knife back.

Set alternative temperature and

wait.

Remove cause.

Replace anti-roll plate.

- - -

Clean.

Cryostat non-operational

Power plug not properly connected.

Defective fuses, or circuit breaker has

triggered.

Check if properly connected.

Replace fuses, or switch circuit

breaker back on. If not possible, call

technical service.

Specimen disc cannot be

removed

Moisture on the underside causes

the specimen to freeze to the freezing

shelf or specimen head.

Apply concentrated alcohol to the

contact point.