Leica CM1950 – Cryostat

8.

Working with the Instrument

• Place working materials such as the blade box or knives (in the knife case), brush, forceps

or preparation needles and, where applicable, specimen discs into the cryostat chamber.

• The necessary tools and preparation aids

can be precooled on the (optional) movable

shelf, making them available at all times in a

convenient position for the user.

• Additionally, specimen discs can be pre-

cooled and stored in the storage system (

see

Chap. 6.5.2 - Fig. 21

).

8.2 Switching on the instrument

The circuit breaker is used simultaneously as the power switch. The switch

must be in the top position for switching on and in the bottom position for

switching off. The switch must be accessible without obstruction.
• Close the sliding window.

The knives are extremely sharp! Handle with care!

Never try to catch a falling knife!

8.1 Preparing cutting tools, specimen discs and preparation aids

For installation of knife/blade holder

and installation in the chamber

see

Chap. 11 "Optional Accessories"

.

The instrument must be switched on at least 5 hours before the

planned use.

To avoid frost formation always put the cover on the quick freeze

shelf.

Always cover the quick freeze shelf during breaks and overnight.