Cleaning the knife holder CN

Daily cleaning

For daily cleaning, it is enough to fold the anti-roll

guide system forwards and remove the section

waste from the knife holder using a dry brush.

Please use a cold brush, as otherwise the sec-

tion waste will thaw and stick to the knife holder.

Fig. 72

1. Fold the anti-roll guide system (33) to the left

while holding it down by the lever (34).

2. Unscrew the bolts (35) of the pressure plate.

3. Afterwards, the pressure plate (36) can be re-

moved for cleaning (with alcohol or acetone).

11.

Optional Accessories

Disinfection (e.g. with Leica Cryofect)

Spray contaminated surfaces with a uniform coat of concentrate or wipe

them down with a rag soaked in it, allow it to make contact for 15 minutes,

then wipe it off.

33

34

35

35

36

For disinfection, commercially avail-

able, mild cleaners and disinfectants

can be used – we recommend Leica

Cryofect.

It is not necessary to oil the parts, such

as the T-piece on the microtome base

plate, clamping lever etc.

The safety gloves included in the stan-

dard scope of delivery must be worn

when cleaning the blade holder to

prevent frostbite of the skin.

Cleaning the blade holder CE

Daily cleaning

The use of flammable sprays within the cryostat chamber is permit-

ted only with the Leica Cryofect tested by us.

Fig. 71