Page 28: Installation

Advertising
background image

28

Instructions for Use V 1.5 RevF - 08/2014

6.5.2 Installing the storage systems (optional)

6.5.3 Shelf, movable (optional)

Attach the rod for the shelf to the inner front

side of the cryostat housing using the provided

screws (1) and the size 3 Allen key, then attach

the caps (3). (The rear side of the movable shelf

has white plastic screws (2) that prevent the

interior of the chamber from being scratched.)

Now hook the movable shelf into the guide rod.

6.5.4 Inserting the section waste tray

6. Installation

To do so, remove the insert (1), place the frame

(2) in front of the bore and tighten the screws/

washer on the cryostat housing using the size 4

Allen key. Afterwards, insert the insert (1) into the

frame and fold it up.

Fig. 23

Before installing the knife/

blade holder base, insert the

section waste tray with the

cutout (a) facing the user.

A

1

2

For reasons of ac-

cessibility, the (op-

tional) storage sys-

tem must always be

installed first.

1

2

3

Fig. 21

Fig. 22