Maintenance and cleaning – Leica Biosystems CM1950 User Manual

Page 74

Advertising
background image

74

Instructions for Use V 1.5 RevF - 08/2014

Removal of the lamp
The UVC lamp (3) is installed in front

of the protection screen for chamber

illumination (4).
• Hold the lamp with both hands and

carefully pull it out of the clips (5)

with a slight forward movement.

• Detach the metal ring (7) on the

holder (6) in the direction of the ar-

row (8) and carefully pull the lamp

out of the holder with your right

hand (see

Fig. 93

)

Installation of the new lamp
• Carefully slide the metal ring (7) over

the lamp from the left (see

Fig. 93

).

• Push the lamp into the holder on the

left side until it engages.

• Screw the metal ring onto the

holder, then hold the lamp with both

hands and carefully push it into the

clips (5).

• Replace the sliding window.
• Connect the instrument to the power

supply again and switch it on.

4

3

5

7

Fig. 92

Fig. 93

Fig. 94

6

8

12. Maintenance and Cleaning

If the UVC key is depressed

for more than 30 seconds,

the running-time meter for

the UVC lamp is reset. This is

required every time the UVC

lamp is replaced to ensure

sufficient power for disinfec-

tion!

Dispose of the UVC lamp separately!