Maintenance and cleaning – Leica Biosystems CM1950 User Manual
Page 74
74
Instructions for Use V 1.5 RevF - 08/2014
Removal of the lamp
The UVC lamp (3) is installed in front
of the protection screen for chamber
illumination (4).
• Hold the lamp with both hands and
carefully pull it out of the clips (5)
with a slight forward movement.
• Detach the metal ring (7) on the
holder (6) in the direction of the ar-
row (8) and carefully pull the lamp
out of the holder with your right
hand (see
Fig. 93
)
Installation of the new lamp
• Carefully slide the metal ring (7) over
the lamp from the left (see
Fig. 93
).
• Push the lamp into the holder on the
left side until it engages.
• Screw the metal ring onto the
holder, then hold the lamp with both
hands and carefully push it into the
clips (5).
• Replace the sliding window.
• Connect the instrument to the power
supply again and switch it on.
4
3
5
7
Fig. 92
Fig. 93
Fig. 94
6
8
12. Maintenance and Cleaning
If the UVC key is depressed
for more than 30 seconds,
the running-time meter for
the UVC lamp is reset. This is
required every time the UVC
lamp is replaced to ensure
sufficient power for disinfec-
tion!
Dispose of the UVC lamp separately!