Leica CM1950 – Cryostat

Installation of the new lamp

• Slightly lift the sliding window (1) holding it by

the grip (2) and pull it out to the front.

1

2

3

4

Removal of the lamp

• The lamp is mounted behind a glare shield and

therefore not visible.

• Touch the lamp for better orientation.
• Lightly tilt the fluorescent tube down to the

right and pull it out of the clip (3).

• Hold the lamp with both hands and pull it to

the right out of the holder (4).

Fig. 97

Fig. 96

Fig. 95

12.

Maintenance and Cleaning

12.3.1 Replacing the fluorescent lamp

Switch the instrument off and discon-

nect the power plug before replacing

the fluorescent lamp!

If the lamp is broken, it must be replaced

by the technical service, as the replace-

ment involves a high risk of injury.

Only use lamps of the same specifica-

tion!

Type:

OSRAM DULUX L 18 W/840.

• Hold the lamp in the correct mounting posi-

tion, as shown, and push it to the right until it

engages in the holder.