1 replacing the fluorescent lamp, Maintenance and cleaning – Leica Biosystems CM1950 User Manual
Page 75
75
Leica CM1950 – Cryostat
Installation of the new lamp
• Slightly lift the sliding window (1) holding it by
the grip (2) and pull it out to the front.
1
2
3
4
Removal of the lamp
• The lamp is mounted behind a glare shield and
therefore not visible.
• Touch the lamp for better orientation.
• Lightly tilt the fluorescent tube down to the
right and pull it out of the clip (3).
• Hold the lamp with both hands and pull it to
the right out of the holder (4).
Fig. 97
Fig. 96
Fig. 95
12.
Maintenance and Cleaning
12.3.1 Replacing the fluorescent lamp
Switch the instrument off and discon-
nect the power plug before replacing
the fluorescent lamp!
If the lamp is broken, it must be replaced
by the technical service, as the replace-
ment involves a high risk of injury.
Only use lamps of the same specifica-
tion!
Type:
OSRAM DULUX L 18 W/840.
• Hold the lamp in the correct mounting posi-
tion, as shown, and push it to the right until it
engages in the holder.