8.

Working with the Instrument

After trimming or sectioning:

Specimen:
• Unclamp and thaw.

• Immerse in fixative for further processing.

Cleaning
• Use the brush to sweep up the section waste (section waste tray) and

remove it from the cryostat (follow applicable laboratory regulations for

disposal).

or
• Clean the cryostat chamber using the chamber suction nozzle:

• To do so, turn the (flat) suction nozzle of the extraction hose by

holding the hose on the white adapter and removing it with a

quick twist. "Park" the flat suction nozzle in a designated place in

the chamber – e.g. on the right inside wall of the cryochamber.

• Remove the chamber suction nozzle from the plastic clip and

attach it firmly to the white adapter.

• Check the remaining capacity of the filter (in

the interior of the chamber) and change the

filter if necessary (

see Chap. 6.5.8

).

• Check the bacteria filter (in the front of the

unit, (

see Chap. 6.5.7

), change it at least every

3 months.

• Move the anti-roll guide to the side (

see

Fig. 61

).

• Spray the cryostat chamber using Cryofect.
• Activate the UVC disinfection.

The use of flammable sprays within the

cryostat chamber

is permitted only with the Leica Cryofect

tested by us.