Working with the Instrument
After trimming or sectioning:
Specimen:
• Unclamp and thaw.
• Immerse in fixative for further processing.
Cleaning
• Use the brush to sweep up the section waste (section waste tray) and
remove it from the cryostat (follow applicable laboratory regulations for
disposal).
or
• Clean the cryostat chamber using the chamber suction nozzle:
• To do so, turn the (flat) suction nozzle of the extraction hose by
holding the hose on the white adapter and removing it with a
quick twist. "Park" the flat suction nozzle in a designated place in
the chamber – e.g. on the right inside wall of the cryochamber.
• Remove the chamber suction nozzle from the plastic clip and
attach it firmly to the white adapter.
• Check the remaining capacity of the filter (in
the interior of the chamber) and change the
filter if necessary (
see Chap. 6.5.8
).
• Check the bacteria filter (in the front of the
unit, (
see Chap. 6.5.7
), change it at least every
3 months.
• Move the anti-roll guide to the side (
see
Fig. 61
).
• Spray the cryostat chamber using Cryofect.
• Activate the UVC disinfection.
The use of flammable sprays within the
cryostat chamber
is permitted only with the Leica Cryofect
tested by us.
Fig. 61