Optional accessories – Leica Biosystems CM1950 User Manual

Page 59

Advertising
background image

59

Leica CM1950 – Cryostat

The knife guard is folded

away in the detail view.

Knife guard folded away

Fig. 65

Four possible settings

Blade rest (straightedge) for low-profile blades

1

2

1

2

6

a

Fig. 66

11.

Optional Accessories

3

4

7a

The anti-roll guide and

pressure plate must be

parallel to each other.

6

• Blade ejector (1) and knife guard (2)

• integrated lateral shift and stable base

• Clearance angle adjustment using size 4 Allen

key (recommended angle between 2° and 5°)

• With finger rest (6a,

Fig. 66

) for brush specimen

• Clamping lever (8) for lateral shift must point

downward to permit unhindered shifting of

finger rest.

• When using high-profile blades, remove the

blade rest (7a,

Fig. 66

).

Conversion of blade holder with anti-roll guide

to blade holder with finger rest
• Screw off the anti-roll guide.

• Unscrew left Allen screw using size 2.5 Allen

key and remove base of anti-roll guide.

• Attach the finger rest (6a) from the left, tighten

the Allen screw using the size 2.5 key – be

careful of the blade ejector!

Blade holder CE with finger rest (for LP + HP)

7

8

5

Red elements on the blade and knife

holders, such as the knife guard and

ejector, are protective devices that may

not be removed.

If you are working with the brush tech-

nique, the knife guard must be folded

upwards.