Optional accessories – Leica Biosystems CM1950 User Manual

Page 64

Advertising
background image

64

Instructions for Use V 1.5 RevF - 08/2014

1

The large specimen disc (80x50 mm) should

preferably be used with knife holder CN and the

16 cm C-profile steel knife.

The clamping jaws are factory-installed in the

knife holder with a clearance of 64 mm. If neces-

sary, both clamping jaws can be offset with a

clearance of 84 mm.

Perform the following steps:
• Use a size 4 Allen key to loosen the screw

over the clearance angle adjustment (1) and

remove the segment arc (2) from the knife

holder base.

• Use a size 4 Allen key to loosen the screws (4)

on the underside of the segment arc.

• Carefully pull the clamping jaw (5) on the right

upward and insert it in the neighboring hole

(6). Tighten the screws from the underside of

the segment arc. Repeat on the left side.

• Insert the long knife support (3) on the side

over the knurled screw (3a) so that the

recess faces the user – turn the height ad-

justment knurled screws until the lower stop

is reached.

• The knife can now be inserted from the side

and its height adjusted via the knurled screws

(3a).

2

3a

4

4

5

6

3

Fig. 73

3a

not

visible

11. Optional Accessories

Never work with only one clamping jaw, as this does not ensure the stability required for the

sectioning process. Also, a long knife will not be sufficiently protected by the knife guard in this

instance.

The specimen disc 50x80 mm is suit-

able only for section thicknesses up to

approx. 5 µm (due to the large specimen

size).

Knife holder CN with anti-roll guide – Moving the clamping jaws and inserting the knife