Working with the instrument – Leica Biosystems CM1950 User Manual

Page 41

Advertising
background image

41

Leica CM1950 – Cryostat

Setting the time
• The clock time is set with the +/- buttons in the control panel marked

with the clock symbol.

To do so, set the current time using the "+" or "-" button below the small

clock symbol.

Pushing the "+" or "-" button for more than 1 sec. increases or decreases

the time continuously (auto-repeat function).

8.

Working with the Instrument

Programming the specimen temperature
• Select the desired temperature of the specimen.

• The specimen temperature is set and indicated on the control panel

marked with the specimen head symbol.

The actual temperature is the standard indication.
Briefly pressing the "+" or "-" button displays the target temperature.
Set the desired value via the "+" / "-" buttons. Pushing the "+" or "-" button

for more than 1 sec. increases or decreases the specimen temperature

continuously.

The actual value will be indicated 5 seconds after finishing the program-

ming.

Specimen temperature - "Max-Cool" function
• The snowflake button for the "Max-Cool" function is in the specimen

temperature field.

Pressing the

button sets the lowest possible specimen head tem-

perature (-50 °C) as the target temperature. The instrument adjusts the

maximum low temperature of the specimen head, i.e. -50 °C.

• Push the snowflake button again to stop the "Max-Cool" function.

The temperature adjusts to the value programmed prior to activating the

"Max-Cool"-function.

• Alternate flashing of "LL" and the actual temperature indicates activation

of the Max-Cool function.

Fig. 43

Fig. 45

Fig. 44

Caution:

The specimen head and Peltier do not switch on until the chamber temperature reaches
-5

°C

, in order to prevent icing.