• Push the knife/blade holder base (2) on the

dovetail guide (4) from the front and tighten it

using the clamping lever (5). Move the clamp-

ing lever clockwise (toward the closed lock

symbol) on the right side of the blade/knife

holder base (

see Detail of Fig. 26

). To move the

base, open the clamping lever only a little to

prevent accidental sliding in the direction of

the specimen head! Move the clamping lever

counterclockwise (toward the open lock sym-

bol) on the right side of the blade/knife holder

base (

see Detail of Fig. 26

).

4

2

6. Installation

A

5

• If the clamping distance is not sufficient, the

clamping lever (3) can be moved. To do so, pull

the lever out and move it to the next position.

Enlarged

detail: 26

When removing the knife holder base (2)

from the refrigerated cryostat chamber,

hold it by the grip points (a – front and

rear) to keep your fingers from freezing.

Safety gloves must be worn!

Fig. 26