2 changing fuses, 3 replacing the uvc lamp, Maintenance and cleaning – Leica Biosystems CM1950 User Manual

Page 73

Advertising
background image

73

Leica CM1950 – Cryostat

12.2 Changing fuses

• In case of power supply faults, please contact an authorized Leica ser-

vice technician immediately.

12.

Maintenance and Cleaning

2

1

Fig. 91

• Switch off the instrument at the circuit breaker.
• Disconnect the power plug.
• Slightly lift the sliding window (1) using the grip

(2) and pull it out to the front.

Turn the instrument off and disconnect the power plug before replacing the UVC lamp.

If the lamp is broken, it must be replaced by the technical service, as the replacement

involves a high risk of injury. Beware of the metallic mercury in the UVC lamp; handle it

carefully and dispose of it properly.

When both LEDs (short and long-term disinfection) are blinking alternately in control

panel field 1, the lamp must be replaced.

A UVC lamp has an estimated service life of approx. 9,000 hours.

Each on/off switching cycle reduces the lamp life by approx. one hour plus

burning time (

30 min

utes or

180 min

utes respectively).

Do not carry out any repairs on your own as this will invalidate the

warranty.

Repairs may only be carried out by qualified service engineers

authorized by Leica.

12.3 Replacing the UVC lamp