Troubleshooting, 1 problems during work, Problems during work – Leica Biosystems CM1950 User Manual

Page 51

Advertising
background image

Leica CM1950 – Cryostat

51

9.

Troubleshooting

9.1 Problems during work

Problem

Cause

Remedy

Frost on chamber walls and

microtome

Cryostat is exposed to air currents

(open windows and doors, air

conditioning).

Frost built-up by breathing into the

cryochamber.

Move to a draft-free location.

Wear mouth protection.

Sections smear

Specimen not cold enough.

Anti-roll plate not yet cold enough,

thus warming the section.

Select lower temperature.

Wait until knife and/or anti-roll

plate have reached chamber

temperature.

Sections splinter

Specimen too cold.

Select higher temperature.

Sections not properly

flattened

Static electricity/air currents.

Specimen not cold enough.

Large area specimen.

Anti-roll plate poorly positioned.

Anti-roll plate poorly aligned with knife

edge.

Incorrect clearance angle.

Blunt knife.

Remove cause.

Select lower temperature.

Trim the specimen parallel, increase

section thickness.

Reposition anti-roll plate.

Align correctly.

Set correct angle.

Use different part of the knife.

Sections not properly

flattened despite correct

temperature and correctly

aligned anti-roll plate

Dirt on knife and/or anti-roll plate.

Edge of anti-roll plate damaged.

Blunt knife.

Clean with dry cloth or brush.

Replace anti-roll plate.

Use different part of the knife.

Sections curl on anti-roll

plate

Anti-roll plate does not protrude far

enough beyond the knife edge.

Readjust correctly.

Scraping noise during

sectioning and specimen

return movement

Anti-roll plate protrudes too far

beyond the knife edge and is scraping

against the specimen.

Readjust correctly.