Troubleshooting
9.1 Problems during work
Problem
Cause
Remedy
Frost on chamber walls and
microtome
Cryostat is exposed to air currents
(open windows and doors, air
conditioning).
Frost built-up by breathing into the
cryochamber.
Move to a draft-free location.
Wear mouth protection.
Sections smear
Specimen not cold enough.
Anti-roll plate not yet cold enough,
thus warming the section.
Select lower temperature.
Wait until knife and/or anti-roll
plate have reached chamber
temperature.
Sections splinter
Specimen too cold.
Select higher temperature.
Sections not properly
flattened
Static electricity/air currents.
Specimen not cold enough.
Large area specimen.
Anti-roll plate poorly positioned.
Anti-roll plate poorly aligned with knife
edge.
Incorrect clearance angle.
Blunt knife.
Remove cause.
Select lower temperature.
Trim the specimen parallel, increase
section thickness.
Reposition anti-roll plate.
Align correctly.
Set correct angle.
Use different part of the knife.
Sections not properly
flattened despite correct
temperature and correctly
aligned anti-roll plate
Dirt on knife and/or anti-roll plate.
Edge of anti-roll plate damaged.
Blunt knife.
Clean with dry cloth or brush.
Replace anti-roll plate.
Use different part of the knife.
Sections curl on anti-roll
plate
Anti-roll plate does not protrude far
enough beyond the knife edge.
Readjust correctly.
Scraping noise during
sectioning and specimen
return movement
Anti-roll plate protrudes too far
beyond the knife edge and is scraping
against the specimen.
Readjust correctly.