Troubleshooting – Leica Biosystems CM1950 User Manual

Page 54

Advertising
background image

54

Instructions for Use V 1.5 - 08/2014

Problem

Cause

Remedy

No or insufficient

refrigeration of the

cryochamber

Cooling system or electronic drive

defective.

Call technical service.

Sliding window collects

condensation

Air humidity and room temperature too

high.

Comply with the requirements for

the installation site.

No or insufficient

refrigeration of the

specimen

Cooling system or electronic drive

defective.

Call technical service.

Sections not properly

flattened

Lamp defective.

Switch defective.

Check lamp and replace it, if

necessary.

Both disinfection LEDs flash

alternately

UV radiation provided by UV tube no

longer sufficient.

Replace UV tube following the

manufacturer’s instructions.

Image of an open-ended wrench

appears due to fault to be rectified

Contact technical service and follow

the instructions given.

9. Troubleshooting