Working with the instrument – Leica Biosystems CM1950 User Manual

Page 44

Advertising
background image

44

Instructions for Use V 1.5 RevF - 08/2014

8.

Working with the Instrument

• Press the TRIM/SECT button. TRIM mode is active if the LED at the top

right is illuminated.

• Set the desired trimming thickness using the "+" or "-" button in control

panel field 2 (for the adjustable sequence of steps,

see Chap. 7 "Instru-

ment Controls" - "Control panel field 2"

).

Entering the section thickness
• Press the TRIM/SECT button. SECT mode is active if the LED at the bot-

tom right is illuminated.

• Set the desired trimming thickness using the "+" or "-" button in control

panel field 2 (for the adjustable sequence of steps,

see Chap. 7 "Instru-

ment Controls" - "Control panel field 2"

).

Switching the retraction on or off in manual sectioning mode
• Press the TRIM/SECT button for approx. 3 sec. The LED in control panel

field 2 displays on

or oFF

.

• You can switch by pressing the "+" or "-" button.
• "Retraction on" means a specimen retraction of 20 µm in manual mode.

During motorized sectioning, the retraction value is speed-dependent and cannot be

changed by the user.

Fig. 51

To switch from a trimming section thickness for research applica-

tions (1- 600 µm) to a thickness for clinical applications (10, 20, 30 or

40 µm), press and hold down the TRIM/SECT key while switching on

the instrument.

Entering the trimming thickness

Fig. 50